Artículo
Signal-to-noise ratio evaluation in resonant ring metamaterial lenses for MRI applications
Autor/es | Algarín Guisado, José Miguel
López Terrones, Marcos Alonso Freire Rosales, Manuel José Marqués Sillero, Ricardo |
Departamento | Universidad de Sevilla. Departamento de Electrónica y Electromagnetismo |
Fecha de publicación | 2011 |
Fecha de depósito | 2017-04-11 |
Publicado en |
|
Resumen | In this paper, we present a method for the evaluation of the signalto- noise ratio in magnetic resonance imaging (MRI) coils loaded with resonant ring metamaterial lenses, in the presence of a conducting phantom resembling ... In this paper, we present a method for the evaluation of the signalto- noise ratio in magnetic resonance imaging (MRI) coils loaded with resonant ring metamaterial lenses, in the presence of a conducting phantom resembling human tissue. The method accounts for the effects of the discrete and finite structure of the metamaterial. Numerical computations are validated with experimental results, including laboratory measurements and MRI experiments. |
Agencias financiadoras | Ministerio de Ciencia e Innovación (MICIN). España Junta de Andalucía |
Identificador del proyecto | CSD2008-00066
TEC2010- 16948 TIC-06238 |
Cita | Algarín Guisado, J.M., López Terrones, M.A., Freire Rosales, M.J. y Marqués Sillero, R. (2011). Signal-to-noise ratio evaluation in resonant ring metamaterial lenses for MRI applications. New Journal of Physics, 13, 115006. |
Ficheros | Tamaño | Formato | Ver | Descripción |
---|---|---|---|---|
Signal to noise.pdf | 1.005Mb | [PDF] | Ver/ | |