Reliability and agreement of three devices for measuring implant stability quotient in the animal ex vivo model
Autor/es | Blázquez-Hinarejos, Mónica
Saka-Herrán, Constanza Díez-Alonso, Víctor Ayuso-Montero, Raúl Velasco-Ortega, Eugenio López López, José |
Departamento | Universidad de Sevilla. Departamento de Estomatología |
Fecha de publicación | 2021-04-12 |
Fecha de depósito | 2021-09-10 |
Publicado en |
|
Resumen | Resonance frequency analysis (RFA) is the most extended method for measuring implant
stability. The implant stability quotient (ISQ) is the measure obtained by different RFA devices;
however, inter- and intra-rater ... Resonance frequency analysis (RFA) is the most extended method for measuring implant stability. The implant stability quotient (ISQ) is the measure obtained by different RFA devices; however, inter- and intra-rater reliability and agreement of these instruments remain unknown. Thirty implants were placed in three different pig mandibles. ISQ was measured parallel and perpendicular (lingual) to the peg axis with Osstell® Beacon, Penguin® and MegaISQ® by two different investigators and furthermore, one performed a test-retest. Intraclass correlation coefficient was calculated to assess the intra- and inter-rater reliability. Pearson correlation coefficient was used to assess the agreement. Intraclass correlation coefficients ranged from 0.20 to 0.65 for the Osstell® Beacon; 0.57 to 0.86 for the Penguin®; and −0.01 to 0.60 for the MegaISQ®. The highest ISQ values were obtained using Penguin® (66.3) in a parallel measurement; the lowest, using the MegaISQ® (60.1) in a parallel measurement. The highest correlation values with the other devices were obtained by MegaISQ® in a parallel measurement. Osstell® Beacon and MegaISQ® showed lower reliability than Penguin®. Osstell® had good agreement for measuring ISQ both in parallel and perpendicular, and MegaISQ® had the best agreement for measuring ISQ in parallel |
Identificador del proyecto | 018582/2019 |
Cita | Blázquez-Hinarejos, M., Saka-Herrán, C., Díez-Alonso, V., Ayuso-Montero, R., Velasco-Ortega, E. y López López, J. (2021). Reliability and agreement of three devices for measuring implant stability quotient in the animal ex vivo model. Applied Sciences, 11 (8). https://doi.org/10.3390/app11083453. |
Ficheros | Tamaño | Formato | Ver | Descripción |
---|---|---|---|---|
Reliability and agreement of ... | 2.597Mb | [PDF] | Ver/ | |