dc.identifier.citationDíaz-Castro, C.M., Falcao, A., López-Jarana, P., Ríos-Santos, J.V., Fernández Palacín, A. y Herrero-Climent, M. (2019). Repeatability of the resonance frequency analysis values in implants with a new technology. Medicina oral, patología oral y cirugía bucal, 24 (5), 22761-E642. https://doi.org/10.4317/medoral.22761.
dc.identifier.issn1698-4447es
dc.identifier.issn1698-6946es
dc.identifier.urihttps://hdl.handle.net/11441/161881
dc.description.abstractBackground: Assess the reliability (by means of reproducibility and repeatability) of the PenguinRFA system, analyse the ISQ values of different implant types and correlate the ISQ with the insertion torque during the place ment of the implant. Material and Methods: 120 rough surface implants were placed in bovine bone (type II and III). The implants were divided into groups, according to its design. Once the implants were in place, the exact insertion torque was regis tered. Then, primary stability was measured by means of the resonance frequency analysis with the PenguinRFA and the Osstell ISQ devices. In each implant two transducers of each device were used. Three measurements were obtained with each transducer. Results: The mean ISQ (implant stability quotient) of the whole sample is 67,70 ± 5,51. The Intraclass Correlation Coefficient (ICC) is 0,933 and 0,944 for transducers 1 and 2 respectively. The reproducibility is 0,906. The mean insertion torque is 24,54 ± 8,96N. The correlation between the ISQ and the insertion torque is 0,507 p<0,000 (MultiPeg 1) and 0,468 p<0,000 (MultiPeg 2) for bone type II and 0,533 p<0,801 (MultiPeg 1) and 0,193 p<0,140 (MultiPeg 2) for bone type III. Conclusions: The results of the present trial suggest that the PenguinRFA presents excellent reproducibility and re peatability, so it could be very useful in the monitoring of the stability of implants over time. Additionally, according to the results, the correlation between the IT and the RFA is low and there are no statistically significant differences in between implant types.es
dc.formatapplication/pdfes
dc.format.extent7 p.es
dc.language.isoenges
dc.publisherMedicina oral S.Les
dc.relation.ispartofMedicina oral, patología oral y cirugía bucal, 24 (5), 22761-E642.
dc.subjectImplant stabilityes
dc.subjectInsertion torquees
dc.subjectISQes
dc.subjectOsseointegrationes
dc.subjectImplant-supported dental prostheseses
dc.subjectImmediate dental implant loadinges
dc.titleRepeatability of the resonance frequency analysis values in implants with a new technologyes
dc.typeinfo:eu-repo/semantics/articlees
dc.type.versioninfo:eu-repo/semantics/publishedVersiones
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.contributor.affiliationUniversidad de Sevilla. Departamento de Estomatologíaes
dc.contributor.affiliationUniversidad de Sevilla. Departamento de Medicina Preventiva y Salud Públicaes
dc.relation.publisherversionhttp://www.medicinaoral.com/medoralfree01/aop/22761.pdfes
dc.identifier.doi10.4317/medoral.22761es
dc.journaltitleMedicina oral, patología oral y cirugía bucales
dc.publication.volumen24es
dc.publication.issue5es
dc.publication.initialPage22761es
dc.publication.endPageE642es

FicherosTamañoFormatoVerDescripción
Repeatability of the resonance ...406.0KbIcon   [PDF] Ver/Abrir  

Este registro aparece en las siguientes colecciones

Mostrar el registro sencillo del ítem

Este documento está protegido por los derechos de propiedad intelectual e industrial. Sin perjuicio de las exenciones legales existentes, queda prohibida su reproducción, distribución, comunicación pública o transformación sin la autorización del titular de los derechos, a menos que se indique lo contrario.