dc.identifier.citationLópez Coto, I., Bolivar, J.P., Más Balbuena, J.L., García-Tenorio García-Balmaseda, R. y Vargas, A. (2007). Development and operational performance of a single calibration chamber for radon detectors. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMEN, 579 (3), 1135-1140. https://doi.org/10.1016/j.nima.2007.06.017.
dc.identifier.issn1872-9576es
dc.identifier.urihttps://hdl.handle.net/11441/156859
dc.description.abstractThis work shows the design, setup and performance of a new single radon detector calibration chamber developed at the University of Huelva (Environmental Radioactivity Group). This system is based on a certified radon source and a traceable reference radon detector, which allows radon concentrations inside the chamber radon to be obtained in steady state conditions within a range of 400-22 000 Bq m-3 18 with associated uncertainties in the range of 4%. In addition, the development of a new ad hoc calibration protocol (UHU-RC/01/06 “Rachel”), which is based on the modelling of radon concentration within the chamber, allows it to be used without the reference detector. To do that, a complete characterization and calibration of the different leakage constants and the flow meter reading have been performed. The accuracy and general performance of both working methods for the same chamber (i.e., with and without the reference detector) have been tested by means of their participation in an intercomparison exercise involving five active radon monitors.es
dc.formatapplication/pdfes
dc.format.extent5es
dc.language.isoenges
dc.publisherElsevier Science BVes
dc.relation.ispartofNUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMEN, 579 (3), 1135-1140.
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectRadon chamber modeles
dc.subjectCalibration protocoles
dc.subjectLeakageses
dc.subjectActive detector inter-comparisones
dc.subjecttraceabilityes
dc.subjectFlow-through radon sourcees
dc.titleDevelopment and operational performance of a single calibration chamber for radon detectorses
dc.typeinfo:eu-repo/semantics/articlees
dc.type.versioninfo:eu-repo/semantics/publishedVersiones
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.contributor.affiliationUniversidad de Sevilla. Departamento de Física Aplicada Ies
dc.identifier.doi10.1016/j.nima.2007.06.017es
dc.journaltitleNUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENes
dc.publication.volumen579es
dc.publication.issue3es
dc.publication.initialPage1135es
dc.publication.endPage1140es

FicherosTamañoFormatoVerDescripción
Rn_Chamber_NIM_A.pdf1.260MbIcon   [PDF] Ver/Abrir  

Este registro aparece en las siguientes colecciones

Mostrar el registro sencillo del ítem

Attribution-NonCommercial-NoDerivatives 4.0 Internacional
Excepto si se señala otra cosa, la licencia del ítem se describe como: Attribution-NonCommercial-NoDerivatives 4.0 Internacional