Por motivos de mantenimiento se ha deshabilitado el inicio de sesión temporalmente. Rogamos disculpen las molestias.
Artículo
Recent developments of the 1 MV AMS facility at the Centro Nacional de Aceleradores
Autor/es | Scognamiglio, Grazia
Chamizo Calvo, Elena López Gutiérrez, José María Müller, A.M. Padilla, S. Santos, F. Javier Vivo Vilches, Carlos García León, Manuel |
Departamento | Universidad de Sevilla. Departamento de Física Aplicada I Universidad de Sevilla. Departamento de Física Atómica, Molecular y Nuclear |
Fecha de publicación | 2016-05 |
Fecha de depósito | 2023-07-18 |
Publicado en |
|
Resumen | The Centro Nacional de Aceleradores (CNA) hosts a 1 MV accelerator mass spectrometry (AMS) apparatus since September 2005. In order to improve its overall performance, several updates have been made on the existing facility ... The Centro Nacional de Aceleradores (CNA) hosts a 1 MV accelerator mass spectrometry (AMS) apparatus since September 2005. In order to improve its overall performance, several updates have been made on the existing facility during the last 10 years of operation. In this paper, two modifications conducted in 2015 will be described. To increase the transmission of the ions through the accelerator, the stripping gas on the 1 MV CNA machine was changed from Ar to He. The measured maximum transmission for almost every isotope results to be higher, especially for heavy masses: for instance, in the case of uranium in the 3+ charge state, the transmission increased from 11% with Ar gas to about 38% with He gas. The second advance consisted of the substitution of the existing gas ionization chamber with a new one provided by ETH Zurich. The ETH detector features with its miniaturized design and is optimized for low energy AMS (i.e. very low electronic noise and efficient charge collection). As the electronic noise is the most important contribution to the resolution for light ions, the total energy resolution has been reduced by 15% in the case of 10Be, allowing a better discrimination against its isobar, 10B. For the heaviest radionuclides where the quality of the spectra is determined by the charge carrier production in the gas, the resolution for 2.7 MeV uranium ions was improved by 30%, probably due to a more efficient charge collection. |
Cita | Scognamiglio, G., Chamizo, E., López Gutiérrez, J.M., Müller, A.M., Padilla, S., Santos, F.J.,...,García León, M. (2016). Recent developments of the 1 MV AMS facility at the Centro Nacional de Aceleradores. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 375, 17-25. https://doi.org/10.1016/j.nimb.2016.03.033. |
Ficheros | Tamaño | Formato | Ver | Descripción |
---|---|---|---|---|
1-s2.0-S0168583X16300088-main.pdf | 1.514Mb | [PDF] | Ver/ | |