Mostrar el registro sencillo del ítem

Artículo

dc.creatorHurtado Bermúdez, Santiago Josées
dc.creatorGarcía León, Manueles
dc.creatorGarcía-Tenorio García-Balmaseda, Rafaeles
dc.date.accessioned2023-05-19T17:41:09Z
dc.date.available2023-05-19T17:41:09Z
dc.date.issued2006
dc.identifier.citationHurtado Bermúdez, S.J., García León, M. y García-Tenorio García-Balmaseda, R. (2006). A revision of energy and resolution calibration method of Ge detectors. Nuclear Instruments and Methods in Physics Research. Section A Accelerators, Spectrometers, Detectors and Associated Equipment, 564 (1), 295-299. https://doi.org/10.1016/j.nima.2006.03.018.
dc.identifier.issn1872-9576es
dc.identifier.issn0168-9002es
dc.identifier.urihttps://hdl.handle.net/11441/146470
dc.description.abstractThe energy and resolution calibration of germanium detectors being basic procedures in gamma-ray spectrometry have not been usually studied in depth. In this paper a new approach to these calibration methods based on the Orthogonal Distance Regression (ODR) algorithm is presented. This algorithm treats more adequately all uncertainties involved in the calibration process and produces better results than the usual Ordinary Least Squares (OLS) algorithm. Additionally a study of the functions involved in energy and resolution calibration of germanium detectors has been also carried out.es
dc.formatapplication/pdfes
dc.format.extent5 p.es
dc.language.isoenges
dc.publisherElsevieres
dc.relation.ispartofNuclear Instruments and Methods in Physics Research. Section A Accelerators, Spectrometers, Detectors and Associated Equipment, 564 (1), 295-299.
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectGamma-spectrometryes
dc.subjectEnergy calibrationes
dc.subjectResolution calibrationes
dc.titleA revision of energy and resolution calibration method of Ge detectorses
dc.typeinfo:eu-repo/semantics/articlees
dcterms.identifierhttps://ror.org/03yxnpp24
dc.type.versioninfo:eu-repo/semantics/publishedVersiones
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.contributor.affiliationUniversidad de Sevilla. Departamento de Física Aplicada IIes
dc.contributor.affiliationUniversidad de Sevilla. Departamento de Física Atómica, Molecular y Nucleares
dc.relation.publisherversionhttps://doi.org/10.1016/j.nima.2006.03.018es
dc.identifier.doi10.1016/j.nima.2006.03.018es
dc.journaltitleNuclear Instruments and Methods in Physics Research. Section A Accelerators, Spectrometers, Detectors and Associated Equipmentes
dc.publication.volumen564es
dc.publication.issue1es
dc.publication.initialPage295es
dc.publication.endPage299es

FicherosTamañoFormatoVerDescripción
1-s2.0-S0168900206005456-main.pdf151.8KbIcon   [PDF] Ver/Abrir  

Este registro aparece en las siguientes colecciones

Mostrar el registro sencillo del ítem

Attribution-NonCommercial-NoDerivatives 4.0 Internacional
Excepto si se señala otra cosa, la licencia del ítem se describe como: Attribution-NonCommercial-NoDerivatives 4.0 Internacional