Article
Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6′)-Ib-cr
Author/s | Calvo, Jorge
Cano, María Elicier Pitart, Cristina Marco, Francesc Rodríguez Martínez, José Manuel Pascual Hernández, Álvaro Martínez Martínez, Luis |
Department | Universidad de Sevilla. Departamento de Microbiología |
Publication Date | 2011 |
Deposit Date | 2017-07-27 |
Published in |
|
Abstract | The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6′)-Ib-cr was evaluated using ... The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6′)-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted. |
Project ID. | PI050690
REIPI RD06/0008 P07-CTS-02908 PI-0282-2010 |
Citation | Calvo, J., Cano, M.E., Pitart, C., Marco, F., Rodríguez Martínez, J.M., Pascual Hernández, Á. y Martínez Martínez, L. (2011). Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6′)-Ib-cr. Journal of Clinical Microbiology, 49 (9), 3343-3345. |
Files | Size | Format | View | Description |
---|---|---|---|---|
Evaluation of Three.pdf | 45.79Kb | [PDF] | View/ | |