García León, José LuisGarcía León, ManuelManjón Collado, Guillermo2025-04-222025-04-222025-08García León, J.L., García León, M. y Manjón Collado, G. (2025). Comparison of the performance of the Quantulus GCT 6220 and the Quantulus 1220 for the determination of 55Fe and 63Ni. Radiation Physics and Chemistry, 233, 112704. https://doi.org/10.1016/j.radphyschem.2025.112704.0969-806X1879-0895https://hdl.handle.net/11441/171909An experimental study of the response of the newer Quantulus GTC 6220 liquid scintillation spectrometer has been conducted with the objective of determining the performance of this counter for the determination of 55Fe and 63Ni in the environment. Its response has been compared to that of the well-established Quantulus 1220. For this, Guard Compensation Technology (GCT) was used to reduce the Lower Limit of Detection (LLD in what follows). Two scintillation cocktails, Hidex ProSafe HC+ and Revvity OptiPhase HiSafe 3 were compared in this work. The parameters employed for this assessment were background count rate, counting efficiency, and LLD.application/pdf6 p.engAttribution-NonCommercial 4.0 Internationalhttp://creativecommons.org/licenses/by-nc/4.0/Quantulus GTC 6220Guard Compensation Technology (GCT)Comparison of the performance of the Quantulus GCT 6220 and the Quantulus 1220 for the determination of 55Fe and 63Niinfo:eu-repo/semantics/articleinfo:eu-repo/semantics/openAccesshttps://doi.org/10.1016/j.radphyschem.2025.112704