2023-05-192023-05-192006Hurtado Bermúdez, S.J., García León, M. y García-Tenorio García-Balmaseda, R. (2006). A revision of energy and resolution calibration method of Ge detectors. Nuclear Instruments and Methods in Physics Research. Section A Accelerators, Spectrometers, Detectors and Associated Equipment, 564 (1), 295-299. https://doi.org/10.1016/j.nima.2006.03.018.1872-95760168-9002https://hdl.handle.net/11441/146470The energy and resolution calibration of germanium detectors being basic procedures in gamma-ray spectrometry have not been usually studied in depth. In this paper a new approach to these calibration methods based on the Orthogonal Distance Regression (ODR) algorithm is presented. This algorithm treats more adequately all uncertainties involved in the calibration process and produces better results than the usual Ordinary Least Squares (OLS) algorithm. Additionally a study of the functions involved in energy and resolution calibration of germanium detectors has been also carried out.application/pdf5 p.engAttribution-NonCommercial-NoDerivatives 4.0 Internacionalhttp://creativecommons.org/licenses/by-nc-nd/4.0/Gamma-spectrometryEnergy calibrationResolution calibrationA revision of energy and resolution calibration method of Ge detectorsinfo:eu-repo/semantics/articleinfo:eu-repo/semantics/openAccess10.1016/j.nima.2006.03.018