Artículo
A simple and fast Double-Flow microfluidic device based liquid-phase microextraction (DF-µLPME) for the determination of parabens in water samples
Autor/es | Ramos Payán, María Dolores
Maspoch, Santiago Llobera, Andreu |
Departamento | Universidad de Sevilla. Departamento de Química Analítica |
Fecha de publicación | 2017-04-01 |
Fecha de depósito | 2024-08-22 |
Publicado en |
|
Resumen | A fast double-flow microfluidic based liquid phase microextraction (DF-µLPME) combined with a HPLC-UV procedure using diode array detection has been developed for the determination of the four most widely used parabens: ... A fast double-flow microfluidic based liquid phase microextraction (DF-µLPME) combined with a HPLC-UV procedure using diode array detection has been developed for the determination of the four most widely used parabens: Ethyl 4-hydroxybenzoate (EtP), Propyl 4-hydroxybenzoate (PrP), Butyl 4-hydroxybenzoate (BuP) and IsoButyl 4-hydroxybenzoate (iBuP) in water samples. Parabens have successfully been determined in environmental (lake and river water) samples with excellent clean up, high extraction efficiency and good enrichment factor using double-flow conditions. The microfluidic device consists of two micro-channels, which contain the acceptor and sample solution separated by a flat membrane (support liquid membrane). The sample (0.32 mM HCl) and acceptor phase (5.6 mM NaOH) are delivered to the µLPME at 10 µL min−1 and 1 µL min−1 flow rate, respectively. The extraction efficiencies are over 84% for all compounds in water samples with enrichment factors within the range of 9–11 and recoveries over 80%. The procedure provides very low detection limits between 1.6 and 3.5 µg L−1. The extraction time and the volume required for the extraction are 5 min and 50 µL, respectively; which are greatly lower compared to any previous extraction procedure for parabens analysis. In addition, this miniaturized DF- µLPME procedure significantly reduces costs compared to not only the existing methods for paraben detection, but also to the existing analytical techniques for sample preparation. |
Cita | Ramos Payán, M.D., Maspoch, S. y Llobera, A. (2017). A simple and fast Double-Flow microfluidic device based liquid-phase microextraction (DF-µLPME) for the determination of parabens in water samples. Talanta, 165, 496-501. https://doi.org/10.1016/j.talanta.2016.12.059. |
Ficheros | Tamaño | Formato | Ver | Descripción |
---|---|---|---|---|
10.1016j.talanta.2016.12.059.pdf | 614.0Kb | [PDF] | Ver/ | Versión aceptada |